C.B.S. Scientific TTGEK-2401-220 User Manual

Page 39

Advertising
background image

TTGE Instruction Manual, version 8/30/2011

39

www.cbsscientific.com

3.6 Gel Casting Techniques- continued

C. Vertical gradient gel casting using gradient maker (GM-40) and a mini-pump for DGGE.

1. Alternatively, you may choose to use a “mini-pump” or other peristaltic pump to cast gels as

shown in (Fig 3-5). If so, secure the gradient maker to a ring stand and connect the outlet

tubing to the mini-pump tubing adapter. Connect tubing from mini-pump to a 20ga. needle for

affixing between glass plates.

Fig. 3-5

2. For TTGE and SSCP single percentage gels, replace the gradient maker with flask (more

consistent than syringe).

Refer to section 3.7 for Gradient Gel Formula Protocol.

See section 4.3 for running conditions.