Step – C.B.S. Scientific TTGEK-2401-220 User Manual

Page 29

Advertising
background image

TTGE Instruction Manual, version 8/30/2011

29

www.cbsscientific.com

Step

Step

Step

Step

Step

3.3 Heater/Stirrer Programming- con’t.

22

Example B
Initial Bath Temperature = 25°C

Program Priority = Temperature

Desired Profile: Cool bath temperature to 20°C and hold it there for 10 minutes.

Decrease

bath

temperature to 10°C over 15 minutes.

Hold bath temperature at 10°C for 15 minutes.

Increase bath temperature to 50°C over a 1-hour period.

This example requires a 5-step program:

Program

Step

Step Set

Point

Step

Duration Controller Operation

1

20°C

10 minutes

Controller cools fluid to 20°C as fast as possible. Until

20°C temperature set point is achieved, “Starting”

appears on the display. When fluid temperature reaches

20°C, “Running” appears on the display.
Fluid temperature is maintained at 20°C for 10 minutes.

2

20°C

15 minutes

Controller cools fluid to 10°C set point programmed for

Step 3. Ramp rate is based on the 15 minute step

duration.

3

10°C

15 minutes Fluid temperature is maintained at 10°C for 15 minutes.

4

10°C

1 hour

Controller slowly heats fluid until 50°C set point

programmed for Step 5 is reached. Ramp rate is based

on the 1 hour step duration.

5

50°C

1 second

“Complete” appears on the display. 50°C fluid

temperature is maintained until set point is changed.

0

5

10

15

20

25

30

35

40

45

50

55

0

5

15

30 45

60

75 90 105 120 135 150 165

Minutes

Degrees C

Step 1

Step 2

Step 3

Step 4

Step 5

Advertising
This manual is related to the following products: