3.3 Heater/Stirrer Programming – continued

3.3.5 Controller Messages

Message Display

Description

Action Required

Standby mode

Normal — Indicates that the Circuit Breaker/Power Switch

is ON and the Controller Power Switch is OFF

An attempt has been made

to set the temperature set

point higher than the Limit

High setting

Error — Decrease temperature set point or increase Limit

High setting

Fluid temperature is higher

than the Alarm High setting

High Temperature Warning — Decrease temperature set

point, increase Alarm High setting, or correct condition

causing high fluid temperature. Pump, heater, and

compressor operation continue.

Fluid temperature is lower

than the Alarm Low setting

Low Temperature Warning— Increase temperature set

point, decrease Alarm Low setting, or correct condition

causing low fluid temperature. Pump, heater, and

compressor operation continue.

An attempt has been made

to set the temperature set

point lower than the Limit

Low setting

Error — Increase temperature set point or decrease Limit

Low setting

Safety Set temperature

exceeded

Safety Fault — Power to heater, compressor, and pump

automatically disconnected. Correct problem and then

press Safety Set Reset Button and ESC Button to clear

fault message and restore operation.

Fluid temperature is higher

than the Limit High setting

Safety Fault — Power to heater and compressor

automatically disconnected. Decrease fluid temperature

or increase Limit High setting. Pump will continue to run.

Fluid temperature is lower

than the Limit Low setting

Safety Fault — Power to heater and compressor

automatically disconnected. Increase fluid temperature or

decrease Limit High setting. Pump will continue to run.

Indicates main power was

lost while a program was

running

Error — Program must be restarted. Press ESC Button to

clear Fault message.

Indicates main power

was lost while a program

paused

Error — Program must be restarted. Press ESC Button to

clear Fault message.

Indicates main power was

lost after a program was

completed, but before it

was exited

Error — Press ESC Button to clear Fault message and

then exit the program.