3.4 Running the Heater/Stirrer

1) Fill reservoir with at least 23 Liters of buffer. (refer to section 3.1 for working buffer volume).

2) Turn Heater/Stirrer “On” (Fig 3-3). Reference section 3.2l Setting the Bath Temperature Set Point (page

15)

a) For DGGE, set at 60ºC.

b) For TGGE, select temperature ramp program, see section 4.3.

Fig. 3-3

3.5

Preparation/Cleaning of Glass Plate for Gel Casting

Hand wash both plates with a high quality lab detergent followed by a complete rinsing with dH

2

O. Air dry or use a

lint-free tissue. Spray/wipe the chosen inner surfaces of the plate set with 95% ethanol and dry with lint-free tissue.