Message Display

Description

Action Required

Analog to Digital

conversion fault

Instrument Failure — Power to heater, compressor, and

pump automatically disconnected. Contact supplier.

Internal temperature probe

fault

Probe Failure — Power to heater, compressor, and pump

automatically disconnected. Contact supplier.

Internal temperature probe

fault

Probe Failure — Power to heater, compressor, and pump

automatically disconnected. Contact supplier.

External temperature

probe fault

Probe Failure — Power to heater, compressor, and

pump automatically disconnected. Replace external

temperature probe or operate instrument using internal

temperature probe. Contact supplier if fault persists.

External temperature

probe fault

Probe Failure — Power to heater, compressor, and

pump automatically disconnected. Replace external

temperature probe or operate instrument using internal

temperature probe. Contact supplier if fault persists.

Internal error

Error — Press ESC Button to continue operation. Contact

supplier.

Triac fault

Triac Failure — Power to heater, compressor, and pump

automatically disconnected. Contact supplier.

Internal memory fault

Failure — Power to heater, compressor, and pump

automatically disconnected. Contact supplier.