TTGE Instruction Manual, version 8/30/2011

21

www.cbsscientific.com

3.2 Heater/Stirrer Set-Up and Programming – continued

To clear an Alarm Low warning, correct the problem or decrease the Alarm Low temperature value.

Limit Low Temperature — This feature provides additional safety and protection by allowing a selectable lower

temperature limit set point.
If you attempt to enter a set point value that exceeds the Limit Low value, the audible alarm will sound and a Warning

message will flash on the display when the Limit Low value is reached. You will also be prevented from decreasing

the set point value any further.

To clear a Limit Low warning, enter a lower value for the Limit High or increase the control temperature set point.
If the Limit Low value is exceeded during operation (due to a Controller fault, excessive cooling load, etc.), a Fault

message will appear on the display and power to the heater and condenser will be disconnected. The pump will

continue to run.
Limits/Alarms Example

Initial Bath Temperature

= 20°C

Bath Temperature Set Point = 37°C

Limit High Temperature

= 60°C

Alarm High Temperature

= 40°C

Alarm Low Temperature

= 35°C

Limit Low Temperature

= 11°C

3.2t Selecting the Pump Speed

Pump speed is selected from the Pump/AutoTune menu. This display shows the current pump speed setting. Five

pump speed settings are available – Maximum, Fast, Medium, Slow, and Minimum.

Select Minimum for DGGE

applications.

If bath temperature moves above

or below

Limit values, alarm

activates and power to the heater

and condenser is disconnected.

Limit Low Set Point. Bath set

point temperature cannot be

lowered below this value.

Low Temperature Alarm disabled

until bath temperature rises above

Low Temperature Alarm set point

value.

High Temperature Alarm Set Point. Alarm activates if bath

temperature rises to this value.

Low Temperature Alarm Set

Point. Alarm activates if bath

temperature falls to this value.

Limit High Set Point. Bath set point

temperature cannot be raised above this value.

10 20 30 40 50 60 70 80 90

M i n u t e s

Degrees