TTGE Instruction Manual, version 8/30/2011

23

www.cbsscientific.com

3.2 Heater/Stirrer Set-Up and Programming – continued

3.2v Selecting the Temperature Probe (Internal or External)

The Probe sub-menu allows you to designate whether to control temperature using the internal bath temperature or

the fluid temperature at an external device. It is available on the Programmable Controller only and requires the use

of an optional external temperature probe.
NOTE: If an external temperature probe is not connected to the Controller, only “Internal” will be available for

selection. For information on connecting an external temperature probe.
When External is selected as the primary temperature probe, the Maximum Setpoint Differential setting becomes

available.

This allows you to set the maximum allowable difference between the set point temperature and the internal bath

temperature. It is intended as a safety feature to protect the internal bath from over-heating or over-cooling in the

event that the external temperature control set point cannot be achieved. The Maximum Setpoint Differential value

may be set from 1° to 10°C. The factory default value is 10°C.

Temperature Control / Display When Using an External Probe

When the difference between the set point temperature and internal bath temperature exceeds the programmed

Maximum Setpoint Differential value, heating is controlled using the internal bath temperature. “Internal” is highlighted

on the temperature display.

Control of the bath heating rate will be based on the internal bath temperature until the difference between the set

point and the internal bath temperature is at the Maximum Setpoint Differential value. When the heating rate is being

controlled using the external bath temperature, the word “External” is highlighted on the display.

NOTE: When the external temperature probe is in use, external bath temperature is displayed in large numerals,

regardless of whether temperature control is based on the internal or external bath temperature.