3.3 Heater/Stirrer Programming- con’t.

21

Programming Examples
Example A
Initial Bath Temperature = 25°C

Program Priority = Temperature
Desired Profile: Cool bath temperature to 20°C and hold it there for 25 minutes.

Increase bath temperature to 30°C and hold it there for 15 minutes.

Increase bath temperature to 50°C over a 45 minutes period.

Decrease bath temperature to 5°C and hold.

This example requires a 6-step program:

Program

Step

Step Set

Point

Step

Duration Controller Operation

1

20°C

25 minutes

Controller cools fluid to 20°C as fast as possible. Until

20°C temperature set point is achieved, “Starting”

appears on the display. When fluid temperature reaches

20°C, “Running” appears on the display.
Fluid temperature maintained at 20°C for 25 minutes.

2

20°C

1 second

Controller heats fluid as fast as possible until the 30°C set

point programmed for Step 3 is reached.

3

30°C

15 minutes Fluid temperature is maintained at 30°C for 15 minutes.

4

30°C

45 minutes

Controller slowly heats fluid until 50°C set point

programmed for Step 5 is reached. Ramp rate is based

on the 45 minute step duration.

5

50°C

1 second

Controller then cools fluid as fast as possible until 5°C set

point programmed for Step 6 is reached.

6

5°C

1 second

“Complete” appears on display. 5°C fluid temperature is

maintained until set point is changed.

0

5

10

15

20

25

30

35

40

45

50

55

0

5

15

30 45

60

75 90 105 120 135 150 165

Minutes

Degrees C