5 Installation

17

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

Ecot

ec E3

000,

k

in

a

22en

1

-r

, 1

407

5

Installation

5.1

Setup

Danger due to moisture and electricity

Moisture getting into the device can lead to personal injury from electric shocks and to
material damage from short circuits.

Operate the Ecotec E3000 only in a dry environment.

Operate the Ecotec E3000 away from sources of liquid and moisture.

Danger due to dropping heavy loads

The device is heavy and can damage persons and items through tilting or dropping.

Place the device only on a sufficiently sturdy surface.

Material damage due to vibration.

Parts of the measurement technology rotate and must not be shaken. The parts continue
to rotate for several minutes after the device is shut down.

Place the device only on a sturdy, vibration-free surface.

The device must not be shaken during operation and at least five minutes after being
switched off.

Material damage due to an overheated device

The device heats up during operation and can overheat without sufficient ventilation.

Please note the specifications,

see page 14

.

Ensure sufficient ventilation especially on the ventilation openings on the left and
right of the device: free space on the side at least 20 cm, in front and in the rear at least
10 cm.

Keep heat sources away from the device.

Do not subject the device to direct sunlight.