Requirements

Measurement requires:

 A sniffer line connected to the main unit.
 The device has run up and warmed up,

see “6.1 Start-up”, page 23

.

 The device is calibrated,

see “6.4.2 Calibration”, page 37

.

 You have made the device settings requires for measurement,

see “6.3 Settings before

measurements”, page 28

.

 You have made the measurement settings requires for measurement,

see “6.4 Settings

for the measurements”, page 34

.

Measuring position and
speed

Hold the sniffer probe to the possible leak as close as possible. The tip may even touch the
test specimen. If a weldseam or the like is to be tested, the tip should be moved along the
path at a speed of less than 10 cm/s. Please also note the minimum measurement time for
the search for helium,

see “Special features with helium”, page 53

.

Test sequence

1 Keep the sniffer probe away from possible gas sources and press the left key on the

sniffer handle (ZERO).

2 Sniff the test specimen.

If there is a leak, it will be reported in the displays, with flashing LEDs in the sniffer handle
and - depending on your settings - also with an acoustic signal.

Danger of eye damage

LEDs generate a bundled light that can damage eyes.

►

Do not look into the LEDs from a short distance or for a longer period of time.

Electric shock hazard

Sucked in liquids can trigger short circuits and cause damage to property or personal
injury.

►

Do not suction in liquids with the device.

►

Use the water conservation tip in humid environments,

see “5.2.1.3 Water

conservation sniffer probe mounting and remove”, page 20

.

Material damage due to a missing sniffer line

The device must not be operated without a connected sniffer line in order to avoid
overpressure in pump and measurement system.

►

Connect the sniffer line before you start up the device.

►

Do not replace the sniffer line while the device is in operation.