W79 Factor outside the range!

An invalid factor was detected during IGS calibration. Repeat the IGS-calibration,

see “IGS alignment”,

page 42

.

If the problem still exists: Contact the INFICON customer
service.

W80 Cathode switched over!

The device was switched over to another cathode
without calibration.

Recalibrate the device,

see page 37

.

W81 Calibration factor too small!

During the internal calibration, the calibration factor
was determined to be < 0.1 or < 0.01 during the
external calibration.

Check the entry for the leak rate,

see “6.4.2 Calibration”,

page 37

.

It was incorrectly calibrated.

Recalibrate the device,

see page 37

.

The leak rate of the calibrated leak is incorrect
(especially during external calibration).

Replace the external calibrated leak.

W82 Calibration factor too large!

During the internal calibration, the calibration factor
was determined to be >10 or > 99.9 during the
external calibration.

Check the entry for the leak rate,

see “6.4.2 Calibration”,

page 37

.

It was incorrectly calibrated.

Recalibrate the device,

see page 37

.

The leak rate of the calibrated leak is incorrect
(especially during external calibration).

Replace the external calibrated leak.

The Transpector sensor is not sensitive.

Contact the INFICON customer service.

E83 Baseline of Transpector not found

The Transpector has a defect.

Contact the INFICON customer service.

E84 Test leak signal too small

(Flow of test leak absolutely too small,
approx. 1 g/a with R134a)

The calibrated leak is too small.

Use a calibrated leak with a higher leak rate.

The calibrated leak is defective.

Check the leak rate of the calibrated leak.

– The Transpector sensitivity is too low.

– The gas flow to the sensor is blocked, the flow

divider is clogged.

Contact the INFICON customer service.

W85 Problem during peak finding!

The Transpector cannot find the peak.
Calibration impossible.

Replace the calibrated leak. If the error still exists contact
the INFICON customer service.

W86 Internal calibration impossible

You are trying to calibrate while the device is not
ready to measure.

– Wait for calibration until the device has run up.

– Call up the measurement view with the menu.

W87 Unsupported gas in the ECO-Check!

– EEPROM in ECO-Check not programmed.

– The ECO-Check is defective.

Use another ECO-Check.

E90 Overspeed (of TMP, E001)

The rotational speed of the turbo molecular pump is
too high.

Contact the INFICON customer service.

E91 Overvoltage (on TMP, E002)

Error in the electric power supply of the turbo
molecular pump

Contact the INFICON customer service.

E92 Run-up time error (E006)

The turbo molecular pump does not run up correctly. Contact the INFICON customer service.

E93 Connection electronic drive unit - pump

defective (E008)

The connection between the control and the turbo
molecular pump is faulty.

Contact the INFICON customer service.

E94 Error in the TC control unit (E015)

Defective turbo molecular pump control.

– Switch the device off. Wait until the pump has

stopped (> 5 min). Switch the device back on.

– Contact the INFICON customer service.

E95 Electronic drive unit does not recognize

pump. (E021)

Error in the communication between turbo
molecular pump and control.

Contact the INFICON customer service.

Table 10: Warning and malfunction messages (cont.)

No. Message

Possible error sources

Fault clearance