2 calibration, Calibration – INFICON Ecotec E3000 Multigas-Sniffer-Leak Detector User Manual

Page 37

Advertising
background image

6 Operation

37

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

Ecot

ec E3

000,

k

in

a

22en

1

-r

, 1

407

Fig. 24: Selection of another mass position

Below the selected mass is the information whether it is the preferred mass or not. In
addition, the molecular mass of the gas as well as the height of the peak in relation to the
highest peak for this gas are displayed. The standard factor is a measure of the sensitivity
of the device for the gas at the set mass position.

Cal. factor

The calibration factor is displayed in this line.

Last cal.

Based on this line, you can check when the last calibration took place.

Cal. method

The line shows whether calibration was external or internal.

6.4.2

Calibration

The Ecotec E3000 can be calibrated most conveniently by adding an ECO-Check calibrated
leak. The ECO-Check can be integrated in the front panel of the device or set up at the
testing position. It compensates temperature fluctuation and thus enables the accuracy
required for calibration.

The ECO-Check calibrated leak includes R134a. It can be used for the calibration of gases
with a mass position from 40 to 105 amu because the Ecotec E3000 converts the
calibration result for the measurement of these gases.

You get the most accurate calibration with the external calibrated leak. The calibrated
leaks are valid for one gas each and temperature-insensitive.

When to calibrate?

The device should be calibrated daily and after every operator change. Calibration is
additionally required after the following events:

 Sniffer line replacement
 Sniffer line probe replacement
 Switch between gases (if you calibrate with an external calibrated leak)
 Filter replacement
 Prompt for calibration by the system

Advertising