7 Warnings and error messages

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

Ecot

ec E3

000,

k

in

a

22en

1

-r

, 1

407

E40 Emission failed

(The emission failed during operation.)

– If the device was switched off for a longer period

of time, this error can occur during the first 10
minutes after start-up.

– The input pressure is too high.

– The Transpector is defective.

Confirm the error message and restart the device. If the
problem still exists: Contact the INFICON customer
service.

E41 No communication with Transpector!

The software cannot establish the connection with
the Transpector.

– Check the connection between Transpector and

wiring board.*

– Contact the INFICON customer service.

The Transpector is defective.

Contact the INFICON customer service.

E42 Transpector temperature > 70 °C or

< 0 °C!

The main air filter is soiled.

Clean or replace the filter,

see page 68

.

Confirm the work,

see page 65

.

– The ambient temperature is too high.

– The ambient temperature is too low.

Please note the ambient conditions,

see page 14

.

E43 Transpector limit value exceeded!

Internal Transpector data error

Contact the INFICON customer service.

E44 Transpector communication error!

Internal Transpector data error

Contact the INFICON customer service.

E45 Transpector hardware error!

Internal Transpector data error

Contact the INFICON customer service.

E46 Transpector hardware warning!

Internal Transpector data error

Contact the INFICON customer service.

E47 Transpector overpressure!

If the device was switched off for a longer period of
time, this error can occur during the first 10 minutes
after start-up.

Confirm the error message and restart the device. If the
problem still exists: Contact the INFICON customer
service.

The sniffer line is not connected.

Connect the sniffer line and confirm the error message.
Restart the device.

E48 Emission failed

(The emission failed during operation.)

– The input pressure is too high.

– The Transpector is defective.

Confirm the error message and restart the device. If the
problem still exists: Contact the INFICON customer
service.

W49 No emission with first cathode

Start-up of the Emission failed. The device has
switched to the second cathode.

You can continue to measure but need to have the
cathodes checked.

E50 Control of turbo pump or electronic

system!

There is an error in the control of the turbo molecular
pump.

Contact the INFICON customer service.

E51 No communication with the turbo

controller!

– Wiring board defective

– CPU board MC50 defective

Contact the INFICON customer service.

W59 Overflow of EEPROM parameter queue! EEPROM defective.

Contact the INFICON customer service.

W60 All EEPROM parameters lost! Please

check your settings!

A new wiring board was installed.

All settings are reset to factory settings. Make your
settings again.

If the message occurs constantly during running up,
the EEPROM on the wiring board is defective.

Contact the INFICON customer service.

W61 0 EEPROM parameters initialized!

Please check the settings!

A software update introduced new parameters. The
new parameters are listed in the warning.

Confirm the warning.

Check whether the factory settings comply with the new
parameters of your application.

If the message occurs constantly during running up,
the EEPROM on the wiring board is defective.

Contact the INFICON customer service.

Table 10: Warning and malfunction messages (cont.)

No. Message

Possible error sources

Fault clearance