7 insert the new porex rods with tweezers – INFICON Ecotec E3000 Multigas-Sniffer-Leak Detector User Manual
Page 70
Advertising
70
8 Maintenance
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
Ecot
ec E3
000,
k
in
a
22en
1
-r
, 1
407
Fig. 48: Fastener of the operating fluid reservoir
4 Lever out the aluminum plug with one or two narrow screw drivers.
5 Hook something into the center bore of the plastic container and pull out the plastic
container.
Fig. 49: Opened operating fluid reservoir
6 Pull the eight Porex rods out of the front side of the bore with tweezers.
Fig. 50: Porex rods of the operating fluid reservoir
7 Insert the new Porex rods with tweezers.