7 Warnings and error messages

59

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

Ecot

ec E3

000,

k

in

a

22en

1

-r

, 1

407

W62 0 EEPROM parameters lost!

Please check the settings!

Changed parameters were detected during the
running up.

The affected parameters are listed in the warning.

Check the settings of the listed parameters.

If the message occurs constantly during running up,
the EEPROM on the wiring board is defective.

Contact the INFICON customer service.

W63 TSP parameters inconsistent! Please

check!

The Transpector was replaced.

– Check the Transpector parameters.

– Contact the INFICON customer service.

The wiring board was replaced.

The EEPROM on the wiring board is defective.

W64 There are warnings pending!

Confirmed but not yet valid warnings are repeated
every two hours or at every new start-up.

– Remedy the cause of the warning.

– Deactivate the warning reminder,

see “Warning

reminder”, page 66

.

W65 TSP serial number inconsistent! Please

check!

The Transpector was replaced.

– Contact the INFICON customer service.

The wiring board was replaced.

The EEPROM on the wiring board is defective.

W66 ECO-Check brand-new!

A new ECO-Check calibrated leak was connected.

Enter the serial number and the code of the calibrated
leak, refer to the installation manual of the calibrated
leak.

W67 ECO-Check expires on DD.MM.YYYY!

The validity of the ECO-Check gas store expires in
three months.

Order an ECO-Check gas store.

W68 ECO-Check expired!

The validity of the ECO-Check gas store is expired (2
years operation or older than 3 years).

Replace the ECO-Check gas store, see installation manual
of the ECO-Check.

W70 All EEPROM parameters of the ECO-Check

lost!

The EEPROM in the ECO-Check calibrated leak is
empty or defective.

Replace the ECO-Check calibrated leak, see installation
manual of the ECO-Check.

W71 No communication with ECO-Check!

The ECO-Check calibrated leak cannot be addressed
by the main unit.

Check the connection to the ECO-Check calibrated leak.
If the problem still exists: Contact the INFICON customer
service.

No ECO-Check calibrated leak connected.

Connect an ECO-Check calibrated leak. If you do not want
to connect the ECO-Check calibrated leak, deactivate the
warning,

see page 34

.

W72 No communication with sniffer line!

The sniffer line cannot be addressed by the main
unit.

Check the connection of sniffer line to main unit
(separate and reconnect; try another sniffer line, if
possible). If the problem still exists, contact the INFICON
customer service.

E73 Unsuitable sniffer line!

The SL3000XL of Protec P3000 was connected
accidentally.

Connect the correct sniffer line.

W77 Changed calibration factor!

(Possible message while the calibration
is being tested.)

The calibration has changed by more than 15 % since
the last calibration changed.

Recalibrate the device,

see page 37

.

W78 Signal difference between test leak and

air too small. (1.25 with R134a)

– The calibrated leak is too small.

– The calibrated leak is defective.

Check the leak rate of the calibrated leak or use a
calibrated leak with a higher leak rate.

The background signal during calibration is too high. Check the background signal by switching off the ZERO

function (press the ZERO key longer than 2 s).

The air signal was not steady during calibration
(confirmed too early).

Recalibrate the device,

see page 37

.

Table 10: Warning and malfunction messages (cont.)

No. Message

Possible

error

sources

Fault clearance