3 replacing mains fuses, Replacing mains fuses – INFICON Ecotec E3000 Multigas-Sniffer-Leak Detector User Manual

Page 71

Advertising
background image

8 Maintenance

71

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

Ecot

ec E3

000,

k

in

a

22en

1

-r

, 1

407

8 Place the plastic container in the bore with the soaked liner first and close it with the

aluminum plug.

9 Tighten the plastic screw again. Ensure that the o-ring is placed correctly in the groove

of the plastic screw and that the bore is lined up properly.

10 Confirm the work,

see “Confirm maintenance”, page 65

.

8.2.3

Replacing mains fuses

The fuses are behind a flap near the mains plug. They are in two slots.

The mains fuses are available with order no. 200 000 914. Two identical fuses must be used
at all times.

1 Lever out the cover of the mains plug to the right with a screwdriver.

Fig. 51: Levering out cover to the right

2 Pull out the two slots and replace the fuses.

Fig. 52: Slot with fuse

3 Push the slots back in. Ensure that the arrows point up.

4 Close the flap.

Danger to life due to electric shock

Inside the device there are voltages of considerable magnitude. Touching parts where
electric voltage is applied can result in death.

►

Disconnect the device from the power supply prior to any maintenance work. Make
sure that the electric power supply cannot be reconnected unauthorized.