5 i&m identification & maintenance functions, 1 i&m0 – INFICON PSG55x ATM to Medium Vacuum Gauge (Profibus) User Manual

Page 21

Advertising
background image

tira56e1 (2010-02) PxG55x.cp

21

5 I&M Identification & Maintenance Functions

I&M is a concept for manufacturer- and sector-independent standardized identi-
fication of field devices. The field device provides specific information in an elec-
tronic nameplate, which can be primarily accessed online with initial operation and
maintenance.
The reading and writing is made on an I&M parameter block consisting of
64 octets. (

→ [4])

Content

Octets

Coding (H)

Header

Manufacturer specific

10

00

H

I&M Block

MANUFACTURER_ID 2

0x017A

H

= INFICON

ORDER_ID 20

Ordering

no.

SERIAL_NUMBER 16

Serial

no.

HARDWARE_REVISION 2

Hardware

revision

1)

SOFTWARE_REVISION 4

Software

revision

1)

REVISION_COUNTER 2

Revision

counter

PROFILE_ID 2

5A00

H

PROFILE_SPECIFIC_TYPE 2

0000

H

IM_VERSION 2

0101

H

IM_SUPPORTED 2

0000

H

1)

of the Profibus board