tira56e1 (2010-02) PxG55x.cp

8 Sensor Analog Input Function Blocks

All gauge functions of the PxG55x are described in the Analog Input Function
Block. Depending on the device configuration there are up to 3 measuring systems
included.

Because the gauge includes three measuring systems, there are three Analog
Input Function Block Instances:
• Instance 1 represents the Capacitance Diaphragm Gauge (CDG) measuring

part of the gauge (

→  34).

• Instance 2 represents the Pirani measuring part of the gauge (→  37).
• Instance 3 represents the Atmosphere Pressure Sensor (ATM) measuring part

of the gauge (

→  40).

Because the gauge includes only one measuring system, there is only one Analog
Input Function Block Instance:
• Instance 1 represents the Pirani measuring part of the gauge (→  37).


This function block describes the functionality of the Capacitance Diaphragm
Gauge (CDG) measuring part of the gauge.

The PSG55x does not have a Capacitance Diaphragm Gauge (CDG)

and this function block (CDG) is not supported.

The following attributes are supported:

ID Name

Structure Data type

Bytes Access

Store

→ 

15 AIBlockAdjust

Command

Record UINT8

2 1_R/W

2_R/W

N 35

16 Block Type ID

Simple

OSTRING(n
)

4 1_R

2_R

N 35

19 ProcessValue

(PV)

Simple *)

– 0_XI

1_R
2_R

D 35

20 Status

Simple UINT8

1 0_XI

1_R
2_R

D 35

21 Data

Type

Simple UINT8

1 2_R/W

N

35

22 Data

Unit

Simple UINT16

2 2_R/W

N

36

23 Reading

Valid

Simple Boolean

1 1_R

2_R

D 36

24 Full

Scale

Simple *)

– 1_R

2_R

N 36

39 Safe

State

Simple UINT8

1 1_R/W

2_R/W

N 36

40 Safe

Value

Simple *)

– 1_R/W

2_R/W

N 36

41 Auto Zero Enable

Simple

Boolean

1

1_R/W
2_R/W

N 36

44 Overrange

Simple *)

– 1_R

2_R

N 36

45 Underrange

Simple *)

– 1_R

2_R

N 36

*) According to data type value (ID 21)

PCG55x

PSG55x

8.1 Sensor Analog Input

Function Block