There are a number of special commands for bringing the gauge into a status it
does not automatically go to.

The device block state method allows the master to force the behavior of the de-
vice block state machine (as shown in parameter device state) by setting it to one
of the values shown in the table below. Depending of the actual state, some values
may not be allowed at a time.

ID value

Name

Description

0 Inactive

No

action.

1

Reset

Used for reinitializing the device.

2

Abort

Brings the device to the ABORT status.

3

Recover

Used for bringing the device from the ABORT
status into the Recovered State ≙ IDLE.

4

Execute

Brings the unit to the EXECUTING status, in
which the gauge functions normally. As soon as
cyclic data traffic is initialized, this status com-
mand is executed automatically.

5

Stop

Brings the gauge to the IDLE status.

6 Perform

Diagnostic

Stops the running activity and starts SELFTEST.

7…127 Reserved

by

PNO

128 Factory

Reset

(Manufacturer
specific)

Used for reinitializing the device (Profibus inter-
face and basic device) to the factory settings.

129 Reset

(Manufacturer
specific)

Used for reinitializing the device (Profibus inter-
face and basic device).

130…255 Manufacturer

specific


7.2.2 Device Block State