INFICON PSG55x ATM to Medium Vacuum Gauge (Profibus) User Manual

Page 28

Advertising
background image

28

tira56e1 (2010-02) PxG55x.cp

The parameter contains an array of 8 bytes, which are assigned as follows:

Byte

no

Name

Description

Value

Common Exception Detail Alarm

0

Size

Number of subsequent bytes
used for description of the alarm
(simple, UINT8, 1 byte)

2

1

Detail 0

Error message
(simple, UINT8, 1 byte)

→ table  28

2

Detail 1

Error message
(simple, UINT8, 1 byte)

→ table  28

Device Exception Detail Alarm

3

Size

Number of subsequent bytes
used for description of the alarm

2

4 Detail

0

5 Detail

1

Error message
(array, UINT8, 2 bytes)

Sensor Alarm (Inst. 1,

ID 103) of PSG55x

Transducer Block

→  50

Manufacturer Exception Detail Alarm

6

Size

Number of subsequent bytes
used for description of the alarm
(simple, UINT8, 1 byte)

1

7 Detail

Error

message

(simple, UINT8, 1 byte)

→ table  28

Bit Common Exception Detail 0

Bit Common Exception Detail 1

0

Internal diagnostic exception

0

0

1 Microprocessor

exception

1

0

2 EPROM

exception

2

0

3 EEPROM

exception

3

0

4 RAM

exception

4

0

5 Communications

exception

5

0

6

Internal real-time exception

6

0

7 0

7 0

Bit Manufacturer Exception Detail

0 Internal

communication

exception

1 Incompatible

software

2 0

3 0

4 0

5 0

6 0

7 0


PSG55x

Common Exception
Detail Alarm / Warning

Manufacturer Exception
Detail Alarm / Warning

Advertising