3 atm transducer block – INFICON PSG55x ATM to Medium Vacuum Gauge (Profibus) User Manual

Page 52

Advertising
background image

52

tira56e1 (2010-02) PxG55x.cp

This transducer block can only be accessed if the ATM sensor is part of the device
configuration (ID 24) (

→  25).

ID Name

Structure Data type

Bytes Access

Store

→ 

101 Block Type ID

Simple

OSTRING(n
)

4 1_R

2_R

N 52

102 Status

Extension

Simple UINT8

1 1_R

2_R

V 52

103 Sensor Alarm

Array

UINT8

2

1_R
2_R

V 52

104 Sensor Warning

Array

UINT8

2

1_R
2_R

V 52

The Block Type ID Parameter contains an ID which describes the block type. The
block type ID of the "ATM Transducer Block" is set to "256". The other defined
block types are listed in Appendix B.

This bit mapped byte parameter provides the following additional information of the
ATM sensor (ATM):

Bit7

Bit6

Bit5

Bit4

Bit3

Bit2

Bit1

Bit0

Res. Res. Res. Res. Res.

Underrange

Exceeded

Overrange

Exceeded

Reading

Invalid

Res. = Reserved
• Underrange

Exceeded

This bit is set ("1") if the ProcessValue PV is below the parameter Underrange
(ID 45) of the ATM sensor instance (

→  42).

• Overrange

Exceeded

This bit is set ("1") if the ProcessValue PV is above the parameter Overrange
(ID 44) of the ATM sensor instance (

→  42).

• Reading

invalid

This bit represents the logical inversion of the parameter "Reading Valid"
(ID 23) of the ATM sensor instance (

→  42).

This parameter contains the following information:

Bit7

Bit6

Bit5

Bit4

Bit3

Bit2

Bit1

Bit0

Byte

0

0 0 0 0 0 0 0 SF

Byte

1

0 0 0 0 0 0 EF 0

• SF = Sensor Element Failure
• EF = Electronics Failure

This byte is always "0". (The ATM Sensor does not produce any warnings).

12.3 ATM Transducer Block

12.3.1 Block Type ID (ID 101)

12.3.2 StatusExtension

(ID 102)

12.3.3 Sensor Alarm (ID 103)

12.3.4 Sensor Warning (ID 104)