3 parameter channel, 1 pke (parameter signature value) – INFICON PSG55x ATM to Medium Vacuum Gauge (Profibus) User Manual

Page 16

Advertising
background image

16

tira56e1 (2010-02) PxG55x.cp

The parameter channel was created as a means to transmit acyclic requests within
a cyclic data stream. Within this section a brief introduction into the matter is
provided. The following table exhibits the principle structure of the parameter chan-
nel.

Octets

Octets

0

1

2

3

4

5

6

7

8

9

10 11 12 13 …

PKE IND res.

PWE

Standard

telegram

Parameter channel (PKW)

PKE

= parameter signature value

IND

= index within a slot

res. =

reserved

PWE =

process

value

Standard telegram = cyclic telegram
PKW

= parameter channel (PKE + IND + PWE)

The PKW allows a read and write access to the parameter space of the slave.
Within the PKW mechanism the master formulates an instruction. The slave pro-
cesses the instruction and formulates the response.
Instructions and responses can’t be blocked: Exactly one instruction is transmitted
in one Output Telegram and exactly one response is transmitted in one Input Tele-
gram. Therefore exactly 4 bytes of user data may be transmitted within one in-
struction or one response.

The instructions / responses are coded in the parameter signature word (PKE):

Bit position

15 14 13 12 11 10

9

8

7

6

5

4

3

2

1

0

AK

res.

PNU (Parameter Number, Slot)


Where: Bits Meaning

15 … 12 AK ≙ Instruction / response signature

11 … 8 reserved

7 … 0

define the slot from which data are read or onto which a
value is to be written

In Master

⇒ Slave communication, the AK field contains the instruction signature of

the master.
In Slave

⇒ Master communication, the AK field contains the instruction signature of

the slave.

AK Function

Master

⇒ Slave

(Instruction signature)

AK

normal

Function
Slave

⇒ Master

(Response signature)

AK

error

0 No instruction

0

No response

1 Read parameter value

1

2

11

Transmit parameter value
(word)
Transmit parameter value
(double word)
Transmit parameter value
(byte)

7

1)

2 Write parameter value

(data type: word)

1

Transmit parameter value
(word)

7

1)

3 Write parameter value

(data type: double word)

2

Transmit parameter value
(double word)

7

1)

10 Write parameter value

(data type: byte)

11

Transmit parameter value
(byte)

7

1)

1)

Instruction cannot be executed (error code)

4.3 Parameter Channel

4.3.1 PKE (parameter

signature value)

Instruction signature

Advertising
This manual is related to the following products: