2 pirani transducer block – INFICON PSG55x ATM to Medium Vacuum Gauge (Profibus) User Manual

Page 51

Advertising
background image

tira56e1 (2010-02) PxG55x.cp

51


ID Name

Structure Data type

Bytes Access

Store

→ 

101 Block Type ID

Simple

OSTRING(n)

4

1_R
2_R

N 51

102 Status

Extension

Simple UINT8

1 1_R

2_R

V 51

103 Sensor Alarm

Array

UINT8

2

1_R
2_R

V 51

104 Sensor

Warning

Array UINT8

2 1_R

2_R

V 51

201 Pirani OFF

Simple

UINT8

1

1_R/W
2_R/W

N 51

The Block Type ID Parameter contains an ID which describes the block type. The
block type ID of the "Pirani Transducer Block" is set to "13". The other defined
block types are listed in Appendix B.

This bit mapped byte parameter provides the following additional information of the
PxG55x sensor (PIR):

Bit7

Bit6

Bit5

Bit4

Bit3

Bit2

Bit1

Bit0

Res. Res. Res. Res. Res.

Underrange

Exceeded

Overrange

Exceeded

Reading

Invalid

Res. = Reserved
• Underrange

Exceeded

This bit is set ("1") if the ProcessValue PV is below the parameter Underrange
(ID 45) of the Pirani sensor instance (

→  39).

• Overrange

Exceeded

This bit is set ("1") if the ProcessValue PV is above the parameter Overrange
(ID 44) of the Pirani sensor instance (

→  39).

• Reading

invalid

This bit represents the logical inversion of the parameter "Reading Valid"
(ID 23) of the Pirani sensor instance (

→  39).

This parameter contains the following information:

Bit7

Bit6

Bit5

Bit4

Bit3

Bit2

Bit1

Bit0

Byte

0

0 0 0 0 0 0 0 SF

Byte

1

0 0 0 0 0 0 EF 0

• SF = Sensor Element Failure
• EF = Electronics Failure

This byte is always "0". (The PIR Sensor does not produce any warnings)

This parameter is used to switch the Pirani off or on (by using an automatic or
manual mode) in case of special gases used in the process.

Parameter value

Description

0

Pirani is always on

1

Pirani switches off automatically
(by default at pressures above 12 mbar - depends on the
system configuration)

2

Pirani is always off


12.2 Pirani Transducer

Block

12.2.1 Block Type ID (ID 101)

12.2.2 StatusExtension

(ID 102)

12.2.3 Sensor Alarm (ID 103)

12.2.4 Sensor Warning (ID 104)

12.2.5 Pirani OFF (ID 201)

Advertising