13 OneOfN Vacuum Gauge Transducer Block

The following attributes are supported:

ID

Name

Structure Data type

Bytes Access

Store

→ 

120 OneOfN

Status

Extension

Simple UINT8

1 1_R

2_R

V 53



This parameter indicates whether the overrange or underrange of the gauge is
exceeded.
If the gauge is operated in its overrange or underrange, the corresponding bit and
additionally the bit "Reading Invalid" is set.
If an error occurs, the bit "Reading Invalid" as well as the corresponding error bits
in Device Block (ID 26, 27, 28) are set.

Bit7

Bit6

Bit5

Bit4

Bit3

Bit2

Bit1

Bit0

0 0 0 0 0 Underrange

exceeded

Overrange

exceeded

Reading

invalid


13.1 Information on the

Individual Indices

13.1.1 OneOfN Status

Extension (ID 120)