11

Leica CM1850 – Cryostat

Extended skin contact with cold parts of the instrument may cause frost
bite!

•

The microtome must be entirely dry before reinstallation. Humidity inside
will condense and freeze in the cold cryostat and thus cause malfunc-
tions or damage.

•

Do not use external heaters for drying the cryochamber. This can cause
damage to the cooling system!

•

All components removed from the cryostat must be carefully dried be-
fore they are replaced in the cryochamber

2.7

Maintenance

Replacement of the fuses

•

Turn the instrument off with the automatic mains fuse and pull the mains
plug, before replacing the fuses.

•

Only use fuses of the same specification! For the required values, please
refer to Chapter 3 ‘Technical data‘.

The use of fuses other than specified by the manufacturer may cause
severe damage to the instrument!

Replacement of the lamp

•

Turn the instrument off with the automatic mains fuse and pull the mains
plug, before replacing the lamp.

•

If the lamp is broken, it must be replaced by the technical service, as the
replacement involves a high risk of injury.