Leica CM 1850 – Cryostat

9.3

Maintenance

9.3.1 General maintenance

The microtome is virtually maintenance-free. To
ensure a smooth operation of the instrument over
several years we recommend the following:

Have the instrument inspected by a qualified
service engineer authorized by us

once a year

.

Enter into a service contract at the end of the
warranty period.
For further information, please contact your
local Leica service center.

Clean the instrument

every day

.

Once a week:

Apply a drop of oil to the plastic coupling

(5,

Fig. 26

)

.

Lubricate the specimen cylinder

(1)

:

Push the appropriate coarse feed button to
move the specimen cylinder out to the front
stop position, apply a drop of cryostat oil and
move the specimen cylinder back to the home
position by pressing the appropriate coarse
feed button.

Occasionally, or when required:

Lubricate the clamping piece (T-piece)

(2)

on

the microtome base plate and the clamping
lever

(3)

.

Lubricate the slot cover

(4)

.

To do so, turn the handwheel to place the
specimen head to the uppermost position and
apply some drops of cryostat oil on to the slot
cover; after that place the specimen head to

Fig. 23

Fig. 24

1

2

3

4

9.

Cleaning, disinfection, maintenance