Troubleshooting – Leica Biosystems CM1850 UV User Manual

Page 38

Advertising
background image

38

Instructions for use V 2.6 – 10/2012

No or insufficient refrigeration

Scraping noise at the microtome

- Compressor defective.
- Leak in the cooling system.
- Inappropriate site conditions.

- Ventilation slits of the liquefier

dirty.

- Friction between the slot cover

and the microtome housing.

- Call technical service.
- Call technical service.
- Check site conditions as de-

scribed in section 5.2.

- Clean the ventilation slits as

described in section 10.3.1.

- Apply cryostat oil to the slot

cover and distribute by turning
the handwheel or with a cloth.

Problem

Cause

Remedy

8.

Troubleshooting

Advertising