Cleaning, disinfection, maintenance – Leica Biosystems CM1850 UV User Manual

Page 39

Advertising
background image

39

Leica CM 1850 – Cryostat

9.1

Cleaning

Remove frozen section waste from the cryostat with a cold brush every
day.

Remove the section waste tray for emptying.

Remove the storage shelves and the brush shelf for cleaning.

Remove the sliding window by slightly lifting and pulling it to the front
when closed

(see 9.3.6 'Replacement of the lamp')

.

Do not use organic solvents or any other aggressive substances for
cleaning and disinfecting! Only use the cleaning agents specified
in this instruction manual such as Leica Cryofect (alcohol or com-
mon disinfectants based on alcohol)!

Drain the cleaning liquid through the hose after the prescribed reagent
time is over and collect it in the waste container

(1)

.

Condensation water produced during defrosting collects in the
waste container. Therefore, check the liquid level regularly and
empty the container if necessary.

9.

Cleaning, disinfection, maintenance

Dispose of the waste liquid according to the
waste disposal regulations.

To remove the waste container

(1)

, shut off the

tap

(2)

and unscrew the lid

(3)

.

1

2

3

Fig. 22