7.

Daily use of the instrument

7.1

Specimen freezing

•

Select the appropriate sectioning temperature (cryochamber tempera-
ture) for the sample material (please refer to

7.4 Temperature selection

chart‘).

7.1.1 Quick freeze shelf

The cryochamber has a quick freeze shelf (

5,

Fig. 17) for up to 10 specimen discs.

The temperature of the quick freeze shelf is always lower than the cryochamber
temperature.

•

Cut the specimen to size.

•

Activate the Peltier element

(4)

, if available - it

may take up to 40 seconds until the maximum
refrigeration output is available.

•

Apply enough cryocompound to a specimen
disc at room temperature.

•

Place the specimen on the disc and orient.

•

Place the specimen disc in one of the holes of
the quick freeze shelf and freeze the specimen
at a low temperature.

•

Once the specimen is frozen, insert the
specimen disc in the specimen head

(Fig. 18)

and start sectioning.

7.1.2 Stationary heat extractor

•

Fix the support

(1)

of the heat extractor

(2)

by tightening the 2 screws to

the threaded holes on the left sidewall of the cryochamber and insert the
heat extractor.

•

Lower the heat extractor cylinder on the specimen surface. After ap-
proximately 30 seconds contact, the specimen will be entirely frozen.

•

Once the specimen is frozen, place the heat extractor on the parking
station

(3)

.

1

2

3

4

5