Leica CM1850 – Cryostat

5.1

Connection to mains power

During the start-up of the compressor the nominal voltage must not
drop below the values specified in the ‘Technical data‘.
Please note that the compressor requires a start-up current be-
tween 45 and 50 A.
Therefore, the electric circuit at the installation site must be in-
spected by an electrical engineer to ensure that it meets the re-
quirements for a smooth operation of the instrument.
A constant adequate power supply to the instrument must be en-
sured at all times.
Failure to comply with the above will cause severe damage to the
instrument.

•

The electric circuit at the place of installation must be protected sepa-

rately.

•

Do not connect any other consumers to this electric circuit.

5.2

Prior to operation

•

Before connecting the instrument to the mains power, please check if

the local mains voltage complies with the power rating indicated on the
name plate of the instrument.

•

Place the storage shelves in the cryochamber.

•

Place the section waste tray and brush shelf in the cryochamber.

•

Place the knife holder base on the microtome base plate.

•

Insert the knife holder and clamp it on the base plate For details, please

refer to the separate instruction manual for your knife holder).

•

Open the knife box with the knife and place it in the cryochamber for

precooling.

•

Place all tools needed for specimen preparation in the cryochamber.

•

Close the sliding window.

•

Connect the mains plug to the mains power outlet at the wall.