Cleaning, disinfection, maintenance – Leica Biosystems CM1850 UV User Manual

Page 42

Advertising
background image

42

Instructions for use V 2.6 – 10/2012

the lowest position and apply some drops of cryostat oil on to the slot
cover; distribute the applied oil by turning the handwheel or with a soft
tissue.

Clean the ventilation slits

(5)

of the liquefier on the right side of the instru-

ment with a brush, broom or vacuum cleaner from dust and dirt in the
direction of the fins.

Do not carry out any repairs on your own as this will invalidate the war-
ranty.
Repairs may only be carried out by qualified service engineers autho-
rized by Leica.

The microtome can be removed for thorough cleaning and disin-
fecting, or for extensive drying after a long power failure!

9.3.2 Removal of the microtome

Turn the instrument off and disconnect the mains plug before re-
moving the microtome.
Prior to removing the microtome, place the specimen head to the
lowest position by placing the handle of the handwheel in the low-
est position.
When removing the microtome, the specimen head will rapidly fall
down and might injure the operator‘s hands.
Wear appropriate protective gloves to take the cold microtome out
of the cryochamber!
Extended skin contact with cold parts of the instrument may cause
frost bite!

Slightly lift the sliding window when closed holding it at the grip provided
and pull it out to the front

(Fig. 30)

- see ‘9.3.6 Replacement of the lamp‘.

Remove the accessories in the following order: brush shelf, knife holder,
section waste tray, specimen discs, stationary heat extractor, left stor-
age shelf, right storage shelf.

9.

Cleaning, disinfection, maintenance