Technical data – Leica Biosystems CM1850 UV User Manual

Page 13

Advertising
background image

13

Leica CM 1850 – Cryostat

Please refer to section 5.2 ‘Site require-
ments‘!

3.

Technical data

All CM 1850 types require the following fuses:

F1:

T0,25 A Fa. Schurter, Typ FST; 6,3x32 mm

or T0,25 A Fa. Littlefuse, Typ 313; 6,3x32 mm

F2:

T0,6 A Fa. Schurter, Typ FST; 6,3x32 mm

or T0,6 A

Fa. Littlefuse, Typ 313; 6,3x32 mm

F3:

T1,6 A Fa. Schurter, Typ FST; 6,3x32 mm

or T1,6 A

Fa. Littlefuse, Typ 313; 6,3x32 mm

F4:

T6,25 A Fa. Schurter, Typ FST; 6,3x32 mm

or T6,25 A Fa. Littlefuse, Typ 313; 6,3x32 mm

F5:

T4 A

Fa. Schurter, Typ FST; 6,3x32 mm

or T4

Fa. Littlefuse, Typ 313; 6,3x32 mm

* Refrigerant and compressor oil must only be replaced by qualified, authorized service per-
sonnel!

Microtome

Rotary microtome
Section thickness setting

1 - 60 μm

Specimen feed

25 mm

Vertical stroke

59 mm

Maximum specimen size

55 x 55 mm

Specimen orientation

8° (x-, y-, z-Achse)

Electric coarse feed

slow

200 μm/s

rapid

700 μm/s

Cryocabinet

Dimensions

Width (w/o handwheel)

600 mm

Width (with handwheel)

730 mm

Depth

730 mm

Height

1140 mm

Weight (incl. microtome,
without specimen cooling) approx.135 kg

Fluorescent lamp

50 Hz version:

Osram Dulux S 11 W/21
Color: LUMILUX light white

60 Hz version:

Osram Dulux S 13 W/21
Color: LUMILUX light white

according to IEC-1010, UL 3101

according to CECOMAF

Liquid temperature 45 °C
Evaporation temperature -25 °C

Storage conditions

Temperature:

5 - 55 °C

Relative humidity: < 85 %, non-condensing

Advertising