Cleaning, disinfection, maintenance – Leica Biosystems CM1850 UV User Manual

Page 43

Advertising
background image

43

Leica CM 1850 – Cryostat

9.

Cleaning, disinfection, maintenance

Loosen the screws

(3)

with an Allen key (4 mm).

Do not loosen the screws (7) for removing the
microtome.

Disconnect the coarse feed motor plug

(1)

by

pulling the metal head.

Disconnect the temperature sensor

(2)

from

the microtome.

1

2

3

4

5

7

8

8

Slightly lift the microtome and pull it to the left
to disengage the plastic coupling

(5)

connecting the axes.

Take the microtome out of the cryochamber.

Fig. 25

Fig. 26