Instructions for Use V 1.9 RevC - 06/2013

2. Safety

Knife guard on the knife holder
Each knife holder is equipped with a tightly

mounted knife guard (8, 9). This makes it possible

to cover completely the cutting edge in every knife

or blade position.

Knife holder N/NZ
The knife guard (8) of the knife holder N/NZ can be

easily positioned via the two handles (7) (

Fig. 4

).

To cover the knife edge, push both cover strips

of the knife guard to the center.

Knife holder E/E-TC
The knife guard on knife holder E/E-TC consists

of a red foldaway handle (9). To cover the cutting

edge, fold the knife guard handle (9) upwards as

illustrated in

Fig. 5

.

Fig. 4

7

7

8

9

10

11

9

Knife holder E-TC

Knife holder N

Knife holder E

The clamping levers on the knife hold-

er E are not interchangeable.

The two clamping levers (10, 11) must

remain in the position shown at all

times, as otherwise isolated malfunc-

tions of the knife holder can occur.

Clamping lever for the blade (10) at the

right, clamping lever for the lateral

displacement (11) at the left.