2 possible faults, Possible faults, Troubleshooting – Leica Biosystems RM2245 User Manual

Page 66

Advertising
background image

64

Instructions for Use V 1.9 RevC - 06/2013

7. Troubleshooting

7.2 Possible faults

Problem

Possible cause

Corrective action

1. Thick/thin sections

The sections alternate between

being thick and thin, or there is

chattering in the sections, or the

specimen is torn out of the em-

bedding. In extreme cases, there

are no sections whatsoever.

• Blade, knife holder or orientation

is not clamped properly.

• The specimen is not clamped

properly.

• Reclamp the blade, knife holder

or orientation.

• Check that the cassette is se-

curely clamped in the universal

cassette clamp.

• If the universal cassette clamp

is soiled by wax, clean the UCC

(refer to Chapter

8.1 - "Cleaning

and Maintenance - Universal

cassette clamp"

).

• When using cassettes whose

lid is molded on, check that the

broken edge allows the cassette

to be clamped securely; if neces-

sary, remove burrs or clamp the

cassette into the universal cas-

sette clamp horizontally instead

of vertically.

• If the cassette dimensions are

within the specified

tolerances and the cassette still

cannot be securely clamped into

place, it may be that the univer-

sal cassette clamp is configured

incorrectly or is defective. In this

case, have Technical Service

inspect and reconfigure the

universal cassette clamp.

• When using cassettes, par-

ticularly thin-walled cassettes,

from a company other than

Leica or Surgipath, the cassette

may become deformed or other

clamping problems may arise.

If while trying to clamp the cas-

sette you realize that it is not

securely clamped into place,

an alternative tensioning clamp

must be used.

Advertising