5. Operation

5.1.3 Display and control elements

Three-digit display

This display is located both on the instrument and

on the control panel.

If the SECT LED is lit up, the display shows the

section thickness setting in µm.

If the TRIM LED is lit up, the display shows the

trimming section thickness setting in µm.

The electric coarse feed at two speeds is used for a rapid movement of the

specimen towards and away from the knife.

With the double-arrow buttons, the coarse feed operates at 900 µm/s; with

the single-arrow buttons, it runs at 300 µm/s.

In sectioning mode the coarse drive can be used in STEP mode, or with

continuous feed. The instrument is delivered with continuous feed (standard

configuration).

Setting the section thickness/trimming section thickness

Adjust these settings using the - keys on the control panel.
Section thickness setting range: 0.50 µm - 100 µm

Setting values: from 0.5 µm - 5.0 µm in 0.5 µm increments

from 5.0 µm - 20.0 µm in 1.0 µm increments

from 20.0 µm - 60.0 µm in 5.0 µm increments

from 60.0 µm - 100.0 µm in 10.0 µm increments

Trimming section thickness setting range: 1 - 600 µm

Setting values: from 1.0 µm - 10.0 µm in 1.0 µm increments

from 10.0 µm - 20.0 µm in 2.0 µm increments

from 20.0 µm - 50.0 µm in 5.0 µm increments

from 50.0 µm - 100.0 µm in 10.0 µm increments

from 100.0 µm - 600.0 µm in 50.0 µm increments

Selecting the sectioning and trimming mode

To switch between sectioning mode and trimming mode, press the TRIM

SECT button. Whenever the button is pressed, the display toggles between

SECT and TRIM.

In the SECT display, the sectioning thickness in the range from 0.50 to

100.0 µm is shown, and in the TRIM display, the trimming section thickness

between 1.0 and 600 µm is displayed.

Coarse feed functions