Leica RM2245

4. Instrument Setup

22

25

24

21

23

The voltage selector is located above the main

power switch, on the left-rear side of the instru-

ment (

Fig.

7). The voltage setting is displayed in

the viewing window (22).
• Insert a small screwdriver into the cutout (25)

and carefully pry out the insert.

• Remove the voltage selector housing (21) to-

gether with the fuses (23). Remove the voltage

selector block (24) (white) and insert it again

such that the correct local voltage is displayed

in the viewing window (22).

• Reinsert the voltage selector housing with the

block and fuses and push it in until it engages

(audible click).

21

26

27

Checking the voltage

(continued)

Connecting the control panel

The connecting cable (15) of the control panel

is fixed to the microtome. It may not be discon-

nected.

• Insert the plug (14) of the connecting cable

(15) into the connecting jack (16) on the rear

of the control panel.

• To fasten the plug, tighten the two screws (17).