8.2.2 Maintenance instructions

The instrument is basically maintenance-free.
To ensure trouble-free operation of the instrument over a long period of time, the following is

recommended by Leica:

3. Have the instrument inspected on a regular

basis by a qualified service technician autho-

rized by Leica.

The intervals depend on how heavily the in-

strument is used.

1. Thoroughly clean the instrument on a daily

basis.

2. Enter into a service contract at the latest at the

end of the warranty period. For more informa-

tion, please contact your local Leica technical

service center.

We recommend the following inspection intervals depending on the general workload of the

instrument as defined in the three categories listed in the table:

8. Cleaning and Maintenance

Note for inspection
• When the instrument reaches approx. 2 million performed strokes, the

code SEr always appears for approx. 3 seconds in the three-digit display

when the TRIM/SECT button is pressed

(switching between trimming and sectioning mode).

This is a reminder that the instrument must be inspected by an authorized

Leica service technician, regardless of how heavily the instrument has

been used.

Only authorized and qualified service personnel may access the internal compo-

nents of the instrument for service and repair!

Category I

Category II

Number of sections per day: > 8,000 sections per day

< 8,000 sections per day

Workload (hours per day):

> 5 hours per day

< 5 hours per day

Sectioning speed:

Predominantly high sectioning speed

Slow to medium sectioning speed

Specimen material:

Working with soft and hard specimens Predominantly soft specimen material

Maintenance:

Every 12 months

Every 24 months