Leica RM2245

6. Optional Accessories

6.8 Freezer pack

98

99

98

6.7 Tray

The freezer pack consists of the freezer plate (98)

and the insulation jacket (99). It is for cooling (or

keeping cold) paraffin specimens. Up to 35 stan-

dard cassettes can be cooled at the same time.

The specimens require approx. 20 min. for being

cooled from room temperature (approx. 20 °C) to

"sectioning temperature."

The cooling performance depends on the ambient

temperature and the volume of the specimens

being cooled.

The insulation jacket can be placed on the hood

of the microtome in the same manner as the tray,

or it can be placed next to the instrument.

The tray is mounted on the hood of the microtome

so that the small feet on the underside fit into the

cutouts on the hood.

It is for storage of the utensils used during sec-

tioning as well as the sectioned specimens.

When using the freezer pack on the

instrument the tray (Fig. 44) must be

removed.

• Remove the freezer plate from the insulation

jacket and place it in a deep freezer, ideally

overnight (but for at least six hours) at approx.

-23 °C.

• Reassemble the freezer plate and insulation

jacket and place them on the bench or the

instrument.