4 adjusting the clearance angle, Adjusting the clearance angle, Operation – Leica Biosystems RM2245 User Manual

Page 34

Advertising
background image

32

Instructions for Use V 1.9 RevC - 06/2013

5. Operation

• The index marks (0°, 5° and 10°) for adjustment

of the clearance angle (59.1) are located on the

right side of the knife holder (57).

• There is also an index mark (59.2) on the right

side of the knife holder base (51) which serves

as a reference point when adjusting the clear-

ance angle.

• Loosen the screw (58) using an Allen key size 4

(71) until the knife holder (57) can be moved.

• Move the knife holder until the index mark of

the desired clearance angle coincides with

the reference line on the knife holder base.

Example:

Enlarged detail showing a clearance angle

setting of 5°.

5.4 Adjusting the clearance angle

51

58

57

59.1

59.2

Enlarged detail:

Index marks for

adjusting the

clearance angle.

71

Fig. 25

The recommended clearance angle

setting for knife holder E is approx. 5°.

• Hold down the knife holder in this position and

retighten the screw (58) for clamping.

Advertising