Operation – Leica Biosystems RM2245 User Manual

Page 28

Advertising
background image

26

Instructions for Use V 1.9 RevC - 06/2013

Configuring the retraction settings

• To call up the retraction settings, press the MENU MODE and CLEAR

buttons simultaneously.

• The current set value is displayed as a three-digit number in the four-digit

display (e.g. "025" = 25 µm).

• Select the desired retraction value.

The retraction value can be adjusted in increments of 5 µm to a maximum

of 100 µm using the - buttons on the control panel.

• To exit the retraction settings, press MENU MODE.

A retraction movement takes place by the newly selected value after

each section.

• While the specimen is in retraction, the yellow LED on the RETRACT

display (4) lights up.

Switching off the specimen retraction

• To call up the retraction settings, press MENU MODE and CLEAR simul-

taneously.

• To switch off retraction, press the button on the control panel until

the display indicates "OFF".

• To exit the retraction settings, press MENU MODE.

When retraction is switched off, the specimen is not retracted.

The yellow LED (4) of the RETRACT indicator does not light up.

5. Operation

+

or

4

+

4

Advertising