7. Troubleshooting

7.1.2 Malfunctions, possible causes and troubleshooting

The following is a list of the most common problems which can arise while working

with the instrument, along with possible causes and troubleshooting procedures.

If the malfunction cannot be remedied with any of the options listed in the table, or the

problem occurs repeatedly, inform Leica Service Support immediately.

Problem

Possible cause

Corrective action

• No display; no reaction to but-

tons pressed after the instru-

ment is switched on.

1. The power cable is not con-

nected correctly.

2. Power fuses are defective.

3. The cable of the control panel

is not connected correctly.

4. The voltage selection switch is

set incorrectly.

1. Check the connection of the

power cable.

2. Replace power fuses (see

Chapter 8.3

).

3. Check the connection of the

cable to the control panel.

4. Check the voltage setting and

correct it if necessary

(see

Chapter 4.4

)

• Feed motion no longer occurs. 1. The end of the residual path

has been reached.

2. The specimen was already

within the residual path when

the instrument was switched

on.

1. Select the trim mode by press-

ing TRIM/SECT. Then return

the specimen using the coarse

feed button. Then retract the

specimen using the coarse

feed button.

2. Follow the procedure in Item 1.

• The specimen in the residual

path area cannot be returned

using the coarse feed buttons.

1. STEP mode is still active (LED

in SECT button lights up).

1. Press the TRIM/SECT button

to switch to trimming mode.

Return the specimen using the

coarse feed button.