Optional accessories – Leica Biosystems RM2245 User Manual

Page 59

Advertising
background image

57

Leica RM2245

2. Attach vertical column

• Insert the cheese head screw (b) into the hole

of the baseplate from below. Place the lock

washer (c) on the cap screw from above.

• Thread the silver vertical column (2) onto the

baseplate from above and tighten with the

Allen key size 8.

A

3. Attach horizontal arm

• Slide the support ring (e) onto the vertical

column and position it so that the lock nut (f)

is facing to the back. Tighten the lock nut.

• Slide the cross-member (d) onto the column.

Ensure that the lock screw (g) is facing to the

right of the baseplate. The horizontal arm must

be centered over the microscope.

• Slide the horizontal arm (3), flat side facing the

lock screw (g), into the cross-member (d) and

tighten.

6. Optional Accessories

Fig. 62

Fig. 61

Fig. 60

b

c

2

e

d

2

g

f

h

The lock washer must be positioned

between the baseplate and vertical col-

umn to prevent unintentional turning of

the column.

Extreme danger of tipping! After install-

ing the vertical column, immediately

place the microtome onto the baseplate

so that the front feet of the microtome

are positioned in the shallow recess-

es (A).

For more information on connecting

and using the microscope, magnifying

lens or cold-light source, please see the

appropriate Instructions for Use.