3 technical data, Technical data, Instrument components and specifications – Leica Biosystems RM2245 User Manual

Page 15

Advertising
background image

13

Leica RM2245

3.3 Technical Data

General

Approvals:

The instrument-specific marks are located next to the

identification label.

Nominal supply voltages:

100/120/230/240 V AC ±10 %

Nominal frequency:

50/60 Hz

Maximum power consumption:

70 VA

Protection class

: I

Power fuses:

2 x T 1.0 A UL listed

Pollution degree

: 2

Overvoltage category

: II

Maximum heat emission:

70 J/s

Operating temperature range:

+10 °C to +35 °C

Temperature range during storage:

+5 °C to +55 °C

Relative humidity:

max. 80 % non-condensing

Storage humidity:

< 80 %

according to IEC-1010, UL 3101, EN 61010

Dimensions and weight
Basic instrument

Width (including handwheel):

413 mm

Width (excluding handwheel):

300 mm

Depth (including waste tray):

618 mm

Height (total):

305 mm (with tray on the hood)

Working height (knife blade):

100 mm (measured from the base plate)

Working height (knife blade):

168 mm (measured from the table)

Weight (without accessories):

approx. 37 kg

Control panel

Width:

94 mm

Depth:

164 mm

Height:

50 mm

Height (in inclined position):

81 mm

Weight (net):

approx. 0.450 kg

3. Instrument Components and Specifications