Instrument setup, 1 standard delivery, 2 installation site requirements – Leica Biosystems RM2245 User Manual

Page 17: Standard delivery, Installation site requirements

Advertising
background image

15

Leica RM2245

4.1 Standard delivery

The Leica RM2245 standard delivery includes:

1 Leica RM2245 basic instrument, without orientation ................................................ 14 0501 38178

1 handwheel, complete .................................................................................................... 14 0501 38181

1 external control panel .................................................................................................... 14 0501 38179

1 section waste tray .......................................................................................................... 14 0502 37931

1 set of power cables consisting of:

1 power cable for Germany ........................................................................................ 14 0411 36958

1 power cable for USA/Canada/Japan ..................................................................... 14 0411 36960

1 power cable for UK ST/BU F-5A .............................................................................. 14 0411 36959

1 tool set - consisting of: ................................................................................................... 14 0502 37965

1 Allen key with handle No. 5 ..................................................................................... 14 0194 04760

1 Allen key with handle No. 4 ..................................................................................... 14 0194 04782

1 Allen key No. 3 .......................................................................................................... 14 0222 04138

1 screwdriver 3x50, 186 long ...................................................................................... 14 0170 11568

1 bottle (50 ml) of oil for drives, type 405 ................................................................... 14 0336 06086

2 micro-fuses 1.0 AT ..................................................................................................... 14 6943 01001

1 brush w/magnet ......................................................................................................... 14 0183 40426

1 dust protective cover ...................................................................................................... 14 0212 30350

1 Instructions for Use ( + language CD) ......................................................................... 14 0501 80101

4. Instrument Setup

4.2 Installation site requirements

• Stable, vibration-free laboratory bench with horizontal, even stage plate;

practically vibration-free floor.

• No other instruments nearby which might cause vibrations.

• Room temperature consistently between +10 °C and +35 °C.

• Obstruction-free access to the handwheel.

The accessories ordered are included in a separate box.

Carefully check the delivery against the packing list and the delivery note. Should there

be any discrepancy, please contact the Leica selling unit handling your order.

Never operate the instrument in rooms with an explosion hazard.