Pressing and holding this key while starting the Pumping Program will start the prime/purge mode. Purge
will begin after the key is held for one second, and continue until the key is released. The pump will stop
after the key is released.

6.11

Dispense Selection

When in Dispense Mode, pressing and holding either the Rate or Volume keys will enter “Dispense

Selection Mode”. The display will show:

nn

, where ‘nn’ is the current dispense selection.

To change the current Dispense Selection, press the up arrow keys below the Dispense Selection’s digits.
The maximum Dispense number is 40. To reset to Dispense number 1, press and hold either the Rate or
Volume keys until Dispense number is 1.

Momentarily press the Rate or Volume keys to exit Dispense Selection Mode.

6.12

‘Program Phase #’ (Number) Key

When in the “Program Entry Mode”, momentary presses of the ’Program Phase #’ and the ‘Program
Function’ keys switch between the Program Phase number and the Program Function displays. The

Program Phase number will be displayed as

nn

, where ‘nn’ is the current Program Phase

number.

When the Program Phase number is displayed and the current Phase’s function is a rate function, a
momentary press of the ‘Program Phase #’ key exits ‘Program Entry Mode, and displays the pumping rate.

To change the current Program Phase number, press the up arrow keys below the Phase number’s digits.
The maximum Phase number is 40. To reset to Phase number 1, press and hold the ‘Program Phase #’ key
until the Phase number is 1.

When a new Program Phase number is selected, the current value of all settings will be that of the
currently selected Program Phase.

6.13

‘Program Function’ Key

When in the “Program Entry Mode”, momentary presses of the ’Program Phase #’ and the ‘Program
Function’ keys switch between the Program Phase number and the Program Function displays.

With the Program Phase function displayed, the Program Function can be selected. Pressing any arrow
key, or an arrow key to the left of the colon (:) or decimal point (.) if displayed with the function, will
select the next Program Function. The selected function is stored by either pressing any non-arrow key, or
after a 2 second delay. If the selected function is different than the original function, the display will blink
when the selected function is stored.

6.13.1

Program Phase Function Parameter

If the selected function has a parameter associated with the function, the value of the parameter will be
displayed to the right of the function name, separated by either a period (.) or a colon (:).

To change the parameter’s value, press the arrow keys below the parameter’s digits. The parameter’s new
value is stored by either pressing any non-arrow key or after a 2 second delay. If the parameter has
changed from its original value, the display will blink when the parameter’s new value is stored.

6.14

‘Setup’ Key

The secondary function of the ‘Diameter’ key is ‘Setup’. While the Pumping Program is not operating,
press and hold the ‘Diameter’ key until the first setup configuration parameter, “Slow Down Mode”, is

displayed:.