Keypad beep enable, Rs-232 pump network configuration, Set tubing inside diameter setting – New Era NE-9000 Series Programmable Peristaltic User Manual

Page 31

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-9000

Publication #1200-04

23

The "Lockout Disable Key" connects the "Program Input", TTL connector pin 6, to Ground, pin 9. The
"Lockout Disable Key" is available as an accessory item.

Lockout can also be disabled by performing a system reset, see sec. 6.17.2, "Reset the Pump."

10.11

Keypad Beep Enable

n

Setting: ‘0’ = Disabled, ‘1’ = Enabled.

When enabled, a single short beep will sound whenever a key is pressed on the keypad.

10.12

RS-232 Pump Network Configuration

nn

The pump can be configured to communicate either with a

computer or another pump. Communications with a computer (Address Mode) is the default setting and
will be indicated by the [Ad:nn] display.

When in the default Address Mode, up to 100 pumps can be attached to a computer in a single pump
network. The network address is defined by the 2 digits to the right of the colon (:). The valid range of
addresses is from ‘00’ to ‘99’. If only one pump is attached to the computer, set the network address to 0,
[Ad:00] (factory default).

After the network address is displayed, the baud rate is displayed. Each pump in the pump network and
the computer must have the same baud rate setting. Any arrow key can be used to scroll through the
selection of baud rates. The supported baud rates are: 300, 1200, 2400, 9600, and 19200 (displayed as
[1920]).

To change the communications mode:

With the network address displayed [Ad:nn], press the left-most arrow key to enter the communications
mode menu. Pressing any arrow key will scroll through the menu selections:

Address Mode: Default communications with a computer mode.

Reciprocating Pumps. Sends Start/Stop, Pumping rate and reverse
pumping direction to an attached secondary pump.

Reciprocating Pumps. Sends Start/Stop, Pumping rate and same
pumping direction to an attached secondary pump.


Note:

Communications with a computer requires the accessory cable: CBL-PC-PUMP-7.

Communications with a secondary pump requires the accessory cable: CBL-DUAL


In Reciprocating or Dual Pumps Modes, the secondary pump must be left in its default communications
setting of Address Mode, Address 0, and 19,200 baud rate. See the documentation with CBL-DUAL
cable for more detailed information.

10.13

Set Tubing Inside Diameter Setting

For most applications, the default tubing inside diameter does not need to be changed since

the pumping head is optimized for 3/16” inside diameter (ID) tubing.

When selected, the display will show the current diameter and the ‘mm’ units LED will blink. While the
units LED is blinking, use the up arrow keys to enter the new diameter. For example, to enter 1/8” id
tubing, use the up arrow keys to change the display to: