11.4.8

Example 8: Dispensing with Complex Synchronization

:01

Repeat if
Program Input
low

:10

:10

Restart on Event Trigger
or 10 seconds

Event Trigger

W

it

h

d

ra

w

I

n

fu

s

e

TTL Program Ouput

TTL High

TTL Low

Event Trigger

Infuse until
Event Trigger

This example demonstrates a complex interaction with external equipment, such as synchronizing with
another peristaltic or syringe pump. The Program includes a variety of interactions with external
equipment, which demonstrates the various control possibilities of the NE-9000.

The Pumping Program begins by canceling any previous event traps (PH:01) and raising the ‘Program
Output’ TTL line (PH:02). After 5.0 ml has been dispensed at 800 ml/min (PH:03), the ‘Program Output’
TTL line is lowered (PH:04), sending a synchronization signal to another device.

The pump then continues to pump at 800 ml/min (PH:06) until a synchronization signal is received at the
‘Event Trigger’ TTL input, causing the Program to jump to Phase 7 (PH:05).