Error and alarm messages, Status messages – New Era NE-9000 Series Programmable Peristaltic User Manual

Page 23

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-9000

Publication #1200-04

15

6.17.4

Special Setup

Pressing the ‘Diameter’/’Setup’ key while turning on power to the pump will enter the Special Setup
selection menu. Use the up arrow keys (

) to scroll through the selections. Press a non-arrow key or

wait for the time out to select.

Set Tubing Inside Diameter

Set Dispense Mode

Set Program Mode Lockout

Exit without making any selection

6.18

Error and Alarm Messages

If the value entered is beyond the pump's capabilities or is invalid, or an operational problem occurred,
one of the following error or alarm messages will be displayed:

Value entered is ‘Out Of Range’ of the pump’s operational limits.

nn

An out of range error occurred at Pumping Program Phase number ‘nn’, or the value just
entered is out of range. Check the pumping parameters and peristaltic tube diameter.

nn

A Pumping Program error was encountered at Pumping Program Phase number ‘nn’.
The indicated Phase is invalid in the context of the entire Pumping Program.

Key pressed is not currently applicable.

A communications time-out alarm occurred with an attached computer while operating in
the “Safe Communications Mode”. This most likely indicates that the RS-232 cable was
detached or the communication program on the computer has ended without turning off
“Safe Communications Mode”.

n

An error was detected during power up, where ‘n’ indicates the error. If n=1, then the
values stored in the pump’s non-volatile memory were invalid and were reset. If n=2,
then the non-volatile memory may need to be replaced.

Pump settings are locked out from the keypad. The lockout key is needed to change
settings. Lockout can also be reset with the reset function.

6.19

Status Messages

Indicates pumping rate units change mode. The units LED's will also be blinking.

Indicates that the Pumping Program has paused and is waiting for the user to press
‘Start’, or for an external operational trigger, to continue.

Indicates that the pump is busy completing a long operation.

Indicates that the “Volume to be Dispensed” is 0.00, and is turned off. This is the
continuous pumping mode.

Indicates that the pump is purging. Displayed while holding down the 'Start/Stop' key.

nn

Indicates either the Dispense Mode dispense number selection, or a Pumping Program is
waiting for the user to select a sub-program.