New Era NE-9000 Series Programmable Peristaltic User Manual

Page 54

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-9000

Publication #1200-04

46

ANTI-DRIP MODE

DRP [ <on-off> ]

Set/query dispense anti-drip mode. Set enables or disables anti-drip mode.

ALARM SETUP

AL [ <on-off> ]

Set/query alarm setup mode. Set alarm enables or disables alarm buzzer mode.

POWER FAIL SETUP

PF [ <on-off> ]

Set/query Power Failure mode. Set Power Failure enables or disables Power Failure mode.

TTL I/O OPERATIONAL TRIGGER SETUP

TRG [ <trigger setup> ]

Set/Query TTL I/O Operational Trigger input configuration.

TTL I/O Operational Trigger is set to <trigger setup>.

<trigger setup> =>

FT

= Foot switch trigger (falling edge start/stop)

FH

= Foot switch hold (falling edge start, rising edge start)

F2

= Foot switch reverse (rising edge start/stop)

LE

= Level trigger (rising edge start, falling edge stop)

ST

= Start only trigger (falling edge start)

T2

= Start only trigger reversed (rising edge start)

SP

= Stop only trigger (falling edge stop)

P2

= Stop only trigger reversed (rising edge stop)

TTL I/O DIRECTIONAL CONTROL INPUT SETUP

DIN [ 0 | 1 ]

Set/query directional control input setup
Settings => 0 = Falling edge: Dispense, Rising edge Withdraw

Same as "dr:rE" setting from the keypad.
Use this setting with the CBL-TTL-1, reciprocating pump cable, to create a 2
pump continuous dispense system

1 = Falling edge: Withdraw, Rising edge Dispense

Same as "dr:dU" setting from the keypad.
Use this setting with the CBL-TTL-1, reciprocating pump cable, to create a 2
pump dual pumping system

PUMP MOTOR OPERATING TTL OUTPUT CONFIGURATION

ROM [ <on-off> ]

Set/query Pump Motor Operating TTL output configuration (TTL pin 7)

Settings => 0 = Output is logic high only when the pump motor is operating (pumping).

1 = Output is logic high when the pump motor is operating (pumping) or when the

Pumping Program is executing a pause timer.

SET KEYPAD LOCKOUT

LOC [P | D] [ <on-off> ]

Set/query keypad lockout mode. Set keypad lockout disables changing any settings from the
keypad unless the "Lockout Disable Key" is inserted.

P [ <on-off> ]